A = logāā(Iā/I) = ε ā cA: without unit; ε: LĀ·molā»Ā¹Ā·cmā»Ā¹ ; ā: cm ; c: mol/L.
Use when: Monochromatic absorption, homogeneous medium, adapted linear domain.
Intu LabUnderstanding through experienceUnderstand the signal, calibration and determination of a concentration.
A device transforms a physical interaction into a measurable signal. UV-visible and colorimetry exploit absorption; ICP-OES measures atomic emissions, atomic absorption an absorption of free atoms, and XRF an X-ray fluorescence.
A calibration line links signal and concentration in a validated domain. The white, the interferences and the matrix condition the quality of the result: the BeerāLambert law does not describe all these devices indiscriminately.
A = logāā(Iā/I) = ε ā cA: without unit; ε: LĀ·molā»Ā¹Ā·cmā»Ā¹ ; ā: cm ; c: mol/L.
Use when: Monochromatic absorption, homogeneous medium, adapted linear domain.
S = a c + b ; c = (S ā b)/aS: signal; a: sensitivity; b: ordered to lāorigin.
Use when: Interpolation in the calibrated range; control the blank and the matrix.
Measure elemental contents in solutions and prepared solids.
Check the composition and traces of impurities.
Calibrate with multiple concentrations, then measure an unknown within the range. Check the effect of a poorly corrected blank.
This sheet explains the principles and their assumptions. Educational simulation can simplify the process; an industrial model requires suitable data and validation.
Only one correct answer per question. Answer all three questions, then view the answer key.